SYNTHES GMBH HOLDING SLEEVE FOR LARGE LOCKING & VA SCREWS; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Catalog Number 314.281
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply.H10 additional narrative: b3: only event year is known.E1 initial reporter address line 1: (b)(6).E3: reporter is a j&j employee.H3, h4, h6: product code: 314.281.Lot number: 22p3667.Manufacturing site: (b)(6).Release to warehouse date:(b)(6) 2020.A manufacturing record evaluation was performed for the finished device lot and no non-conformance was identified.Visual analysis of the returned sample revealed that the guide shaft (part number: 507899) of the hold-sl f/scr f/lcp 4.5/5 was missing.No other issue was identified.A dimensional inspection for the hold-sl f/scr f/lcp 4.5/5 was not performed as it is not applicable to the complaint condition.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the hold-sl f/scr f/lcp 4.5/5 would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from synthes reports an event in portugal as follows: it was reported that during inspection at local distribution center it was identified that the device is damaged.During manufacturer's investigation of the returned device it was identified that the guide shaft was missing.This report is for one (1) holding sleeve for large locking & va screws.This is report 1 of 1 for complaint (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOLDING SLEEVE FOR LARGE LOCKING & VA SCREWS
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
WERK BETTLACH (CH)
muracherstrasse 3
bettlach 2544
SZ   2544
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key16844739
Report Number8030965-2023-05474
Device Sequence Number6823221
Product Code LXH
UDI-Device Identifier07611819248496
UDI-Public07611819248496
Combination Product (Y/N)N
Initial Reporter CountryPO
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number314.281
Device Lot Number22P3667
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/31/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/27/2023
Initial Report FDA Received Date05/01/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/31/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-