| Model Number |
8881520673 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/03/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported that upon completion of treatment, the tip of the syringe broke off in the patient's hd catheter.The clinician, educator and nephrologist are aware.The tip was finally removed with forceps and no damage was seen to the catheter hub.Per additional information received, there was no medical intervention required.
|
| |
|
Additional Manufacturer Narrative
|
|
An investigation is currently underway.Upon completion, the results will be forwarded.
|
| |
|
Additional Manufacturer Narrative
|
|
A decontaminated sample was received at the plant for the investigation.The device history record was reviewed and indicated that the product was released accomplishing all quality standards.Upon a visual evaluation of the sample, the reported issue was confirmed; the luer tip of the syringe barrel broke off near the base of the syringe barrel face.
|
| |
|