MAUDE Adverse Event Report: ARTHREX, INC. IBAL UKA TIB BEA-RING TRL SZ 4 8MM; ORTHOPEDIC, MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. IBAL UKA TIB BEA-RING TRL SZ 4 8MM; ORTHOPEDIC, MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number IBAL UKA TIB BEA-RING TRL SZ 4 8MM
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/01/2023
Type of Reportable Event Malfunction
Event or Problem Description
On 04/05/2023, it was reported by a sales representative via sems that an ar-601-tbd8 tibial bearing trial broke.This occurred during a case, with no patient effect reported.No additional information was provided.Additional information is requested.
 
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Additional Manufacturer Narrative
The complaint allegation is confirmed based on the customer-provided photo.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to user error of the device due to user-applied mechanical forces during insertion/extraction of the trial.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IBAL UKA TIB BEA-RING TRL SZ 4 8MM
Common Device Name
ORTHOPEDIC, MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
vik bajnath
8009337001
MDR Report Key16846008
Report Number1220246-2023-06648
Device Sequence Number7547663
Product Code HWT
UDI-Device Identifier00888867037625
UDI-Public00888867037625
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 10/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberIBAL UKA TIB BEA-RING TRL SZ 4 8MM
Device Catalogue NumberAR-601-TBD8
Device Lot Number5261910
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/05/2023
Supplement Date Received by Manufacturer10/04/2024
Initial Report FDA Received Date05/01/2023
Supplement Report FDA Received Date10/14/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/21/2019
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-