| Model Number |
2274-08-000 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
04/18/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during case physician was tightening screw.In the process the drill bit broke and was attached to the head of the screw.It took the physician twenty minutes to remove the broken hex drill bit.X rays were performed.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # : (b)(4).Investigation summary : an analysis of the product could not be performed since a physical sample was not received for evaluation.An evaluation of the manufacturing record could not be performed as the required product identification number was not provided to complete the evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.However, if the product or product id numbers are received at a later date, the investigation will be updated as applicable.
|
| |
|
Search Alerts/Recalls
|