| Model Number |
100/597/080 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/14/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Other, other text: investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when required.
|
| |
|
Event or Problem Description
|
|
It was reported that the cannula was leaking at the entrance from the blue cuff lumen into the cannula.Trach change took place regarding defective cuff.No injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Other, other text: one device was returned for investigation in used conditions.Functional testing was done where the cuff was attempted to be inflated.The cuff did not inflate.During the manufacturing procedure, each device cuff is inflated for 12 hours.Due to the cuff issue being found while in use, it is most probable that the issue happened after the device left the manufacturing site.A device history record (dhr) reviewed show no non-conformities during the manufacturing of the reported lot number.No further action was taken.
|
| |
|