• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALLERGAN (COSTA RICA) NATRELLE TE SMOOTH 133S-MX-16-T; EXPANDER, SKIN, INFLATABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALLERGAN (COSTA RICA) NATRELLE TE SMOOTH 133S-MX-16-T; EXPANDER, SKIN, INFLATABLE Back to Search Results
Model Number 133S-MX-16-T
Medical Device Problem Code Deflation Problem (1149)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 04/12/2023
Type of Reportable Event Serious Injury
Event or Problem Description
Healthcare professional reported right side deflation.Device remains implanted.
 
Additional Manufacturer Narrative
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: deflation.
 
Event or Problem Description
Device has been explanted and replaced.
 
Event or Problem Description
Healthcare professional reported right side deflation.Device has been explanted and replaced.
 
Additional Manufacturer Narrative
Device evaluation: based on the device analysis grid, the assessments of the complaint are: deflation: observed, one opening anterior side assessed as surgical damage and one opening device on suture tab edge side assessed as unidentified (tear) opening.Additional observations: crease is observed.Yellow particles in device inner surface are observed.No further actions are required as the device was implanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NATRELLE TE SMOOTH 133S-MX-16-T
Common Device Name
EXPANDER, SKIN, INFLATABLE
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer (Section G)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
Manufacturer Contact
terry ingram
12331-a riata trace parkway
building 3
austin, TX 78727
8479366324
MDR Report Key16850346
Report Number9617229-2023-07522
Device Sequence Number6771361
Product Code LCJ
UDI-Device Identifier10888628044227
UDI-Public10888628044227
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K182054
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number133S-MX-16-T
Device Catalogue Number133S-MX-16-T
Device Lot Number3609790
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/19/2023
Supplement Date Received by Manufacturer05/09/2023
07/05/2023
Initial Report FDA Received Date05/01/2023
Supplement Report FDA Received Date05/25/2023
07/28/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/02/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age62 YR
Patient SexFemale
Patient Weight152 KG
-
-