MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR; DEFIBRILLATION LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR; DEFIBRILLATION LEAD Back to Search Results
Model Number 7121Q/52
Medical Device Problem Codes Over-Sensing (1438); Low impedance (2285)
Health Effect - Clinical Code Low Blood Pressure/ Hypotension (1914)
Date of Event 04/12/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient presented in clinic for right ventricular (rv) lead and right atrial (ra) lead revision.It was previously noted that the rv lead exhibited noise over-sensing and low pacing impedance and the ra lead exhibited inappropriate automatic mode switch due to noise over-sensing.Both leads were capped and replaced on (b)(6) 2023.Patient was experiencing low blood pressure during procedure.The hypotension was resolved, and patient condition was stable post procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR
Common Device Name
DEFIBRILLATION LEAD
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key16850908
Report Number2017865-2023-18910
Device Sequence Number7648076
Product Code NVY
UDI-Device Identifier05414734503211
UDI-Public05414734503211
Combination Product (Y/N)Y
Initial Reporter StateNM
Initial Reporter CountryUS
PMA/510(K) Number
P950022
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2011
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2011
Device Model Number7121Q/52
Device Catalogue Number7121Q-52
Device Lot Number3242744
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/12/2023
Initial Report FDA Received Date05/02/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/28/2010
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
QUADRA ASSURA; QUARTETE
Outcome Attributed to Adverse Event Required Intervention;
Patient Age78 YR
Patient SexMale
Patient Weight84 KG
-
-