DEPUY ORTHOPAEDICS INC US PINNACLE 100 ACET CUP 56MM; PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS Back to Search Results
Model Number 1217-01-056
Medical Device Problem Code Osseointegration Problem (3003)
Health Effect - Clinical Codes Extraskeletal ossification (1428); Fatigue (1849); Hypersensitivity/Allergic reaction (1907); Pain (1994); Osteolysis (2377); Numbness (2415); Ambulation or Postural Difficulties (2544); Inadequate Osseointegration (2646); Metal Related Pathology (4530); Swelling/ Edema (4577)
Date of Event 01/01/2023
Type of Reportable Event Serious Injury
Event or Problem Description
On (b)(6) 2021: patient has a consult regarding his right hip with persistent swelling that has occurred nearly a year after his revision.He has been aspirated several times with no signs of infection.Chromium and cobalt levels are normal.He has some numbness and pain radiating down the back of his leg.Mri shows a very large pseudotumor around the right hip that has recurred as well as possible loosening of the right cup.X-ray review shows osteolysis around the stem proximally on the medial side.On (b)(6) 2022: physical therapy note indicates patient had a second pseudotumor removal from his hip on (b)(6) 2021.He has weakness and difficulty walking requiring physical therapy treatments.It is unknown if any implants were revised.On (b)(6) 2022: patient cancels physical therapy visits due to excessive ¿popping sounds¿ in his hip.Doi: (b)(6) 2020; dor: unknown; right hip second complaint.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).E3 initial reporter occupation: lawyer.Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint (b)(4).Investigation summary: no device associated with this report was received for examination.The product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information, monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added: b5, h6 (clinical code).If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
On (b)(6) 2022, x-ray findings right total hip prosthesis with acetabular cup and femoral stem indicated that there were no significant changes in irregularity and minimal heterotopic ossification involving the right greater trochanter.Surgical drain in place with small amount of pneumarthrosis and subcutaneous emphysema.No evidence of periprosthetic fracture.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINNACLE 100 ACET CUP 56MM
Common Device Name
PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key16852332
Report Number1818910-2023-09307
Device Sequence Number20745193
Product Code LPH
UDI-Device Identifier10603295008576
UDI-Public10603295008576
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
P070026
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2005
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1217-01-056
Device Catalogue Number121701056
Device Lot NumberZ48ET1000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/07/2023
Supplement Date Received by Manufacturer06/06/2023
07/12/2023
Initial Report FDA Received Date05/02/2023
Supplement Report FDA Received Date06/06/2023
07/19/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/21/2005
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ALTRX +4 10D 40IDX56O; DLT TS CER HD 12/14 40MM +8.5; PINNACLE 100 ACET CUP 56MM; SOL SYS 8/18.0 LRG ST
Outcome Attributed to Adverse Event Required Intervention;
Patient Age60 YR
Patient SexMale
-
-