• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. RINGLOC BI-POLAR 28X54MM; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. RINGLOC BI-POLAR 28X54MM; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Medical Device Problem Codes Unstable (1667); Detachment of Device or Device Component (2907); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Fall (1848); Bone Fracture(s) (1870); Failure of Implant (1924); Pain (1994); Joint Dislocation (2374); Ambulation or Postural Difficulties (2544); Osteopenia/ Osteoporosis (2651); Muscle/Tendon Damage (4532); Swelling/ Edema (4577)
Date of Event 02/03/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that patient fell and experienced a dislocation after a bipolar hip procedure.The surgeon tried to reduce but was unable to.It was then found that the head had separated form the liner and bipolar.The patient underwent a revision surgery approximately a week later.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Additional Manufacturer Narrative
(b)(4).D10: 00771101620 ¿ extended offset stem ¿ 63299354.802202803 ¿ femoral head ¿ 65633002.Product will not be returning to zimmer biomet for the investigation as the product location is unknown.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that patient experienced a fall and dislocation approximately 1-2 weeks after initial hip procedure.The surgeon attempted to reduce the dislocation and noticed that the head had disassociated from the liner and bipolar shell.The patient ultimately underwent a revision surgery.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.00771101500 ¿ stem ¿ 64918519.802202802 ¿ cocr head ¿ 65396111.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No additional event information to report at this time.
 
Additional Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.Visual examination of the returned product identified the liner was returned installed inside of the shell.There are scratches and indentations on the liner.The shell and liner could not be separated for further evaluation.Dhr was reviewed and no discrepancies related to the reported event were found.Root cause was unable to be determined.Medical records were provided and reviewed by a health care professional.Review of the available records identified the following: an initial left hemiarthroplasty was performed due to a fall and femoral neck fracture.The patient reported falling at home and dislocating.There was a closed reduction, but the patient dislocated again and many times after.A new fracture on the femur was also found.Later testing showed disassociation of the bipolar components.During the revision, it was confirmed the bipolar disassociated from the head.All components were revised and replaced.The complaint is confirmed based on the provided medical records.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
It was reported that patient experienced a fall and dislocation approximately 3 weeks after a left hemiarthroplasty.A closed reduction was performed, but the patient dislocated again upon attempting to ambulate, closed reduction was performed again and patient was admitted to the hospital.A new greater trochanter fracture was found and determined no surgery to be performed and the patient was discharged to home.Approximately 1 week later, the patient experienced another fall and dislocation with admittance to the hospital.A closed reduction under anesthesia was attempted but unsuccessful.The surgeon noted disassociation of the products along with unstable hip fracture.The next day, the patient underwent a revision surgery of all products.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RINGLOC BI-POLAR 28X54MM
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key16854717
Report Number0001825034-2023-00925
Device Sequence Number15345028
Product Code JDI
UDI-Device Identifier00880304001800
UDI-Public(01)00880304001800(17)251210(10)580550
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K051569
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 09/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/10/2025
Device Model NumberN/A
Device Catalogue Number11-165232
Device Lot Number580550
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/07/2023
Supplement Date Received by Manufacturer05/01/2023
05/25/2023
09/13/2023
Initial Report FDA Received Date05/02/2023
Supplement Report FDA Received Date05/25/2023
06/14/2023
09/15/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/10/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10.
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexMale
Patient Weight84 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-