| Model Number |
U411 |
| Medical Device Problem Code |
Unable to Obtain Readings (1516)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/17/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The field service engineer exchanged the test strip tray and cleaned a test strip sensor.Two runs of patient samples were processed and these were processed correctly.H3 other text : na.
|
| |
|
Event or Problem Description
|
|
The initial reporter stated they had an issue with mismatched data on the cobas u 411 urine analyzer.There appears to be some overlap of data as data measured from a first sample will be associated with a second sample.For example, sample a will run on the analyzer and results will be measured for this sample.Sample b is then run on the analyzer and no results will be measured.Sample c is then run on the analyzer and the results measured from sample b then appear for sample c.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the alarm information from the analyzer determined there was a software error and a measurement inconsistency error.The remedy for both errors requires a re-start of the system.If this is not done, this can lead to a sample mismatch.The system data logs were not available, so it could not be determined if the system was re-started.The field service engineer re-installed the software, verified the test strip sensor area was clean.Calibration and controls were processed successfully and no further issues have arisen.The investigation determined the service actions resolved the issue.
|
| |
|