• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MENTOR TEXAS MENTOR SMOOTH ROUND SPECTRUM; PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MENTOR TEXAS MENTOR SMOOTH ROUND SPECTRUM; PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE Back to Search Results
Model Number 3501485
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Deformity/ Disfigurement (2360)
Date of Event 04/05/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a 34-year-old caucasian female patient underwent a primary breast augmentation with a 525cc mentor smooth round spectrum saline breast prosthesis and experienced a rupture on the right side post-operatively, which was diagnosed during an office visit.As a result, the patient underwent explantation on (b)(6) 2023.
 
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: deflation.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
On may 10, 2023, the mentor failure analysis lab has received the device for evaluation.On may 17, 2023, the evaluation for the device was completed.Device evaluation summary: visual analysis of the returned sample revealed that the spec smooth rnd 525cc breast implant had a crease/fold on the posterior view.Leak testing was performed, according to mentor procedure, and it identified a tear within the crease/fold, measuring approximately 0.2 cm.Additionally, no other leak sites were detected.The evaluation determined that the possible cause of the rupture is consistent with normal wear.Deflation can occur at any time after implantation, but it is more likely to occur the longer the implant is implanted.Creating wrinkles or folds should be avoided during implantation or other procedures as it can result in deflation at a later time.Breast implants may also simply wear out over time.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Capa activity was not required.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post-market surveillance.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR SMOOTH ROUND SPECTRUM
Common Device Name
PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
kate karberg
3041 skyway circle north
irving, TX 75038
3035526892
MDR Report Key16855247
Report Number1645337-2023-05117
Device Sequence Number6526532
Product Code FWM
UDI-Device Identifier00081317002031
UDI-Public00081317002031
Combination Product (Y/N)N
Initial Reporter StateAR
Initial Reporter CountryUS
PMA/510(K) Number
P990075
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2016
Device Model Number3501485
Device Catalogue Number3501485
Device Lot Number6636589
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/10/2023
Supplement Date Received by Manufacturer05/10/2023
Initial Report FDA Received Date05/02/2023
Supplement Report FDA Received Date05/31/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/21/2012
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age34 YR
Patient SexFemale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-