The patient was initially treated for an abdominal aortic aneurysm (aaa) with the implant of an alto abdominal stent graft system (two ovation ix iliac limbs were implanted in the left iliac), on (b)(6) 2022.On (b)(6) 2023, a type iiia endoleak with aneurysm enlargement was identified.Reintervention was completed on (b)(6) 2023.At reintervention the exact endoleak point was not determined.A gore excluder leg (non-endologix) was implanted into each limb.The type iiia endoleak was resolved.The patient was reported to be stable.
|
|
The reported incident was investigated in alignment with endologix operating procedures and work instructions.Where possible, it is endologix practice to make at least three good faith efforts to retrieve a reported incident related device as well as medical records and medical imaging.An evaluation of the manufacturing record was completed.A review of the part number/lot number combination needed for device identification shows the device to have been properly manufactured and released in accordance with the device master record.An evaluation of the device could not be completed.The device was not returned to endologix for evaluation because it remains implanted.A clinical evaluation of the incident could not be completed.No medical records nor medical imaging relevant to the reported incident was received by endologix.The user facility was unable to provide this information.Due to the absence of medical records and medical imaging device, use, procedure, and/or anatomy relatedness to this incident could not be evaluated.The final patient status remains unknown.The final patient status was not made available to endologix.No additional investigation of this reported incident is planned.However, should additional information relevant to the investigation outcome become available, a follow-up report will be submitted.Endologix will continue to monitor this and similar incidents.H3 other text : device remains implanted.
|