Model Number N/A
Medical Device Problem Codes Fracture (1260); Difficult to Insert (1316)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/14/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a total knee arthroplasty was performed.During the operation, the surgeon failed to insert an articular surface several times.He removed it several times and the back side on the articular surface broke.There was no consequences or impact to the patient.Attempts have been made and no further information has been provided.
 
Additional Manufacturer Narrative
(b)(4).Concomitant medical products: 42532007101 - natural tibia cemented 5 degree stemmed left size e - 65740117.Reporting source: foreign - japan.The product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Visual examination of the returned product identified damage to the distal surface (nicked/gouged) and the locking feature (dovetail) is found to be flared out and compressed.The device history record was reviewed and no discrepancies relevant to the reported event were found.Damage to the articular surface can occur if the articular surface is not correctly placed and oriented before pushing it using the inserter instrument.Detailed instructions for inserting the articular surface are provided in the persona the personalized knee surgical technique.The root cause is attributed to use error.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTICULAR SURFACE MEDIAL CONGRUENT (MC) LEFT 12 MM HEIGHT USE WITH TIBIA
Common Device Name
INSTRUMENT, KNEE
Manufacturer (Section D)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI 
Manufacturer (Section G)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI  
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key16857202
Report Number3007963827-2023-00108
Device Sequence Number8555894
Product Code MBH
UDI-Device Identifier00889024469310
UDI-Public(01)00889024469310(17)260512(10)65011264
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K150090
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/12/2026
Device Model NumberN/A
Device Catalogue Number42512100812
Device Lot Number65011264
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/27/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/14/2023
Supplement Date Received by Manufacturer08/17/2023
Initial Report FDA Received Date05/02/2023
Supplement Report FDA Received Date08/17/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/03/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10 NARRATIVE.