| Model Number |
PM3542 |
| Medical Device Problem Code |
Failure to Interrogate (1332)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/21/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the device could not be interrogated.The device was explanted and replaced to resolve the event.The patient was stable and there were no adverse consequences.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event of wireless communication problem was not confirmed.Device image analysis indicated excessive high powered telemetry usage which caused device to temporarily disable rf telemetry to preserve battery power.The monthly current trend was normal, and voltage was in normal range of operation.Further electrical and mechanical analysis performed in nominal setting did not find any anomaly and device was able to be interrogated through rf telemetry.Longevity assessment was performed, and device was found to have appropriate remaining longevity.
|
| |
|