| Model Number |
SSM-560 |
| Medical Device Problem Code |
Device Appears to Trigger Rejection (1524)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture baker grade iv.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side capsular contracture baker grade iv.Device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: based on the device analysis grid, the assessments of the complaint are: ¿ capsular contracture: unable to observe as it is a medical event not related to the device.Additional observations: ¿ crease flat and white particles observed on the device.No further actions are required as the device was implanted.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side capsular contracture baker grade iv.Device has been explanted and replaced.
|
| |
|
Event or Problem Description
|
|
Device has been explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.
|
| |
|