Model Number 482804745
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 04/04/2023
Type of Reportable Event Malfunction
Event or Problem Description
A company representative reported that a patient implanted with es2 pedicle screws for a one level lumbar fusion presented with a surgical infection approximately one-month post-operatively.During removal of the implanted hardware it was discovered that tulip of an es2 long blade cannulated screw had disengaged.The tulip of the screw was explanted but the shaft of the screw was unable to be removed and was left in the pedicle.
 
Additional Manufacturer Narrative
H6 codes have been updated to reflect receipt of the device and conclusion of the investigation.
 
Event or Problem Description
A company representative reported that a patient implanted with es2 pedicle screws for a one level lumbar fusion presented with a surgical infection approximately one-month post-operatively.During removal of the implanted hardware it was discovered that tulip of an es2 long blade cannulated screw had disengaged.The tulip of the screw was explanted but the shaft of the screw was unable to be removed and was left in the pedicle.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ES2 INTEGRATED BLADE SCREW SIZE L 7.5X45MM
Common Device Name
THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM
Manufacturer (Section D)
STRYKER SPINE-US
2 pearl court
allendale NJ 07401
Manufacturer (Section G)
STRYKER SPINE-FRANCE
zone industrielle de marticot
cestas 33610
FR   33610
Manufacturer Contact
rita karan
2 pearl court
allendale, NJ 07401
2017608000
MDR Report Key16857579
Report Number0009617544-2023-00019
Device Sequence Number3428807
Product Code NKB
UDI-Device Identifier07613327002232
UDI-Public07613327002232
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
K122845
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number482804745
Device Catalogue Number482804745
Device Lot Number178045
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/19/2023
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/04/2023
Supplement Date Received by Manufacturer07/26/2023
Initial Report FDA Received Date05/02/2023
Supplement Report FDA Received Date08/22/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/02/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexMale
-
-