H10: as this malfunction is considered one event identified by the dealer; only one mdr report will be submitted for the reported quantity affected of forty for this event.H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: four impra eptfe patches was received for evaluation.Upon visual evaluation, sample one and four was noted to have black specs and yellow discoloration foreign material were noted within inner packaging.Sample two and three was noted to have black specs within inner packaging.No functional performed due to the nature of the complaint.Therefore, the investigation is confirmed for the for the reported contamination issue.A definitive root cause for the alleged contamination issue could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiration date: 06/2027), g3 h11: h6 (method, result, conclusion) h11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
|