• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER ORTHOPAEDIC MFG. LTD. FEMUR CEMENTED POSTERIOR STABILIZED (PS) NARROW RIGHT SIZE 5; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER ORTHOPAEDIC MFG. LTD. FEMUR CEMENTED POSTERIOR STABILIZED (PS) NARROW RIGHT SIZE 5; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Pain (1994); Loss of Range of Motion (2032)
Date of Event 07/15/2021
Type of Reportable Event Serious Injury
Event or Problem Description
 it was reported a patient underwent an initial right total knee arthroplasty performed.Subsequently, one month post procedure, the patient underwent a manipulation under anesthesia due to limited range of motion.No additional complications have been reported.
 
Additional Manufacturer Narrative
(b)(4).Arthrofibrosis is defined as the development of fibrous scar tissue within or surrounding a joint.Arthrofibrosis is a known postoperative procedure related complication that can occur from surgical implantation of new joint replacement as well as from previous injuries or surgical procedures.Scar tissue formation is a normal healing response; however, the buildup of such can result in pain, stiffness, limited range of motion, and difficulty properly ambulating.If excess scar tissue develops, conservative measures such as exercises or physical therapy would be attempted first.If these attempts fail, surgical intervention such as manipulation under anesthesia, arthroscopic arthrolysis, or open arthrotomy would become necessary to remove the fibrous tissue and restore joint function.Additional associated products & mdrs: 42522600410 articular surface fixed bearing (cps) rt 10mm lot# 64623989.Mdr: 3007963827-2023-00105.Additional associated products: 42532006402 tibia cemented 5 degree stemmed right size c lot# 64884533.42540000029 all poly patella cemented 29 mm dia lot# 64843618.Unk palacos r&g cement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FEMUR CEMENTED POSTERIOR STABILIZED (PS) NARROW RIGHT SIZE 5
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI 
Manufacturer (Section G)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI  
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key16859825
Report Number3007963827-2023-00104
Device Sequence Number7262294
Product Code JWH
UDI-Device Identifier00889024228726
UDI-Public(01)00889024228726(17)300817(10)64771556
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K113369
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Study,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number42500005802
Device Lot Number64771556
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/19/2023
Initial Report FDA Received Date05/03/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/24/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient SexFemale
Patient Weight69 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-