| Model Number |
68LP-350 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
03/15/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been/will be initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.Further information from the reporter regarding event, product, or patient details has been/will be requested.No additional information is available at this time.The reason for reoperation is: deflation.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported a left side deflation.Device remains implanted.
|
| |
|
Event or Problem Description
|
|
Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: based on the device analysis grid, the assessments of the complaint are: deflation: one opening observed on the posterior side assessed as fold flaw opening.Additional observation: ¿ crease was observed.¿ no penetrating nick ( stress marks) observed.¿ wear abrasion observed.No further actions are required as no issues with the manufacturing process are observed.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported a left side deflation.Device has been explanted.
|
| |
|
Search Alerts/Recalls
|