Model Number 1218-36-658
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 04/22/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # : (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that patient had multiple surgeries.On (b)(6) 2021 hemi arthroplasty was done and stem remains from this procedure.On 9/19/21 hip dislocated and converted to constrained total hip by surgeon.On 12/14/21 cup was revised and dual mobility was placed by surgeon.Der was filed, cup and screw that listed on the der, placed at this time.On (b)(6) 2022 it was converted to dm to constrained due to dislocation.Der was submitted, head and liner placed at this time.Patient presents in er with a periprosthetic facture of right hip with appearance of possible infection superior to acetabular component.All components removed and prostalac placed uneventfully.There was loosening of femoral component post facture at bone to implant interface.Doi: (b)(6) 2021.Dor: (b)(6) 2023.Affected side: right hip.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Event or Problem Description
Additional information received indicated that there were no instrumentations that broke.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PIN LNR CONS NEUT +4 36IDX58OD
Common Device Name
PINNACLE HIP SYSTEM : HIP CONSTRAINED ACETABULAR LINERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key16859923
Report Number1818910-2023-09460
Device Sequence Number14908624
Product Code KWZ
UDI-Device Identifier10603295012139
UDI-Public10603295012139
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K043058
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number1218-36-658
Device Catalogue Number121836658
Device Lot NumberJD8828
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/22/2023
Supplement Date Received by Manufacturer05/03/2023
05/19/2023
Initial Report FDA Received Date05/03/2023
Supplement Report FDA Received Date05/11/2023
05/21/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/18/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ARTICULEZE M HEAD 36MM +15.5; CORAIL AMT COLLAR SIZE 12; PINN CAN BONE SCREW 6.5MMX20MM; PINN SECTOR W/GRIPTION 58MM; UNKNOWN BONE CEMENT
Outcome Attributed to Adverse Event Required Intervention;
Patient Age86 YR
Patient SexMale
-
-