| Model Number |
38151 |
| Medical Device Problem Code |
Temperature Problem (3022)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/04/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that there was an issue with the product 38151 robi plastic handle.According to the information received the doctor was shocked and burned.There was no patient harm/injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The manufactures internal number is (b)(4).Investigation is on-going.Product will not be returned.Ifu im-000004 v11.0(01-2019) bipolar instruments and take-apart bipolar instruments warns to check the bipolar cord for cuts,crimps or burns and to inspect all bipolar instruments prior to and after each procedure for visiable signs of damage.
|
| |
|
Additional Manufacturer Narrative
|
|
A bipolar cable that was used in the event was left off of section d10 of original filing.
|
| |
|
Additional Manufacturer Narrative
|
|
Determination: since the item is not sent for examination, it cannot be examined more closely.According to the customer description, the doctor has received a shock / burn.A possible cause may be that the doctor wore no, or defective gloves and the current could thus flow in his body or handle was wet, so that the current was passed here.Ifu im-000004 ifu v11.0 01-2019 warns " use surgical gloves specifically designated by their manufacturer for electro-surgical procedures.The use of other gloves may result in electrosurgical shocks or burns to the surgeon.".
|
| |
|