| Model Number |
96-0082 |
| Medical Device Problem Code |
Loss of or Failure to Bond (1068)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
04/21/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Patient was revised due to loosening of femoral component.Adapter bolt was broken at threads.All parts were removed.Sleeve and stem were well fixed and were not removed.S rom hinge femur was implanted.Tibial component was not revised.Doi: unknown.Dor: (b)(6) 2023.Affected side: left knee.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot a manufacturing record evaluation (mre), was not possible because the required lot code was not provided.
|
| |
|
Search Alerts/Recalls
|