ZIMMER MANUFACTURING B.V. ALL-POLY PATELLA CEMENTED 32 MM DIAMETER; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Failure of Implant (1924); Unspecified Infection (1930); Pain (1994); Inadequate Osseointegration (2646)
Date of Event 04/13/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported a patient underwent a left knee arthroplasty.At 3 months post op, the patient was pain free with 120 degrees flexion.Approximately 4 months post implantation, the patient presented with knee pain.The original reason for revision was for suspected infection, however, the surgeon mentioned the knee did not look infected when removing the prosthesis.The surgeon took 6 swabs, 3 on distal femur and 3 on proximal tibia.When removing the tibia, it was noted that the medial tm peg had snapped prior to attempts to remove.The bone ingrowth on the underside of tm tibia was small.The peg that was left in patient had bone ingrowth.It was noted that the medial tm peg is not perpendicular to the baseplate in x-rays.Additionally, metal debris seen in x-rays.Attempts have been made and no further information has been provided.
 
Additional Manufacturer Narrative
(b)(4).G2: foreign - australia.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001822565-2023-01195, 3007963827-2023-00107, 0001822565-2023-01196.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Visual examination of the provided pictures identified sign of being in-vivo.One of the pegs from the tibial plate was fractured and it was removed from the patient.The removed peg shows sign of bone in growth.The device history record was reviewed and no discrepancies relevant to the reported event were found.Medical records/radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: anatomic alignment of the left knee arthroplasty.Radiolucency along the tibial implant as noted.No sign of peg fracture were noted.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALL-POLY PATELLA CEMENTED 32 MM DIAMETER
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER MANUFACTURING B.V.
turpeaux industrial park
route #1 km 123.4 bldg #1
mercedita PR 00715
Manufacturer (Section G)
ZIMMER MANUFACTURING B.V.
turpeaux industrial park
route #1 km 123.4 bldg #1
mercedita PR 00715
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key16860500
Report Number0002648920-2023-00084
Device Sequence Number14904370
Product Code MBH
UDI-Device Identifier00889024247697
UDI-Public(01)00889024247697(17)261114(10)65271072
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K172524
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number42540200032
Device Lot Number65271072
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/13/2023
Supplement Date Received by Manufacturer08/17/2023
Initial Report FDA Received Date05/03/2023
Supplement Report FDA Received Date08/17/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/14/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexFemale
Patient Weight125 KG
-
-