The 2.6f picc was returned for evaluation.Visual inspection confirms the complaint as the lumen burst at the 5 cm mark.The hole created is approximately 0.3 cm in length.A small amount of blood was visible just distal to the hole but was easily flushed out.A definitive root cause cannot be determined.The device subassembly is received from an external supplier and is visually and functionally inspected.Incoming inspection includes visual inspection for kinks, holes, gaps, and functional testing for leaks.At the completion of the final manufacture processes, the catheters are 100% leak tested.Catheter lumens with damage, such as holes, ruptures, tears and/or small pinholes will be detected during this test.Excessive external pressure applied to the lumen may be a contributing cause for this type of failure.The instructions for use (ifu) contains the following warnings and precautions: *small syringes will generate excessive pressure and may damage the catheter.The use of 10cc or larger syringes are recommended.*do not use sharp instruments near the extension lines or catheter lumen.*do not use scissors to remove dressing.*catheter will be damaged if clamps other than what is provided with this kit are used.*clamping of the tubing repeatedly in the same location will weaken tubing.Avoid clamping near the luer(s) and hub of the catheter.
|