COVIDIEN MFG DC BOULDER FORCETRIAD; ELECTROSURGICAL, CUTTING & COAGULATION & ACCES Back to Search Results
Model Number FORCETRIAD
Medical Device Problem Code Energy Output Problem (1431)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/13/2023
Type of Reportable Event Malfunction
Event or Problem Description
According to the reporter, during lap pancreatectomy, there were good seals halfway through then the device was not sealing, there was no re-grasp alert and no evidence of any energy delivery but there was an end tone, there was a usual number of cleaning the jaw.Changed handpiece and generator then the sealing was okay.There was no patient injury.
 
Additional Manufacturer Narrative
D10 concomitant product: lf1937, jaw lap lf1937 ligasure maryland 37cm (lot#: 30380099x).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FORCETRIAD
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCES
Manufacturer (Section D)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
Manufacturer (Section G)
COVIDIEN MFG DC BOULDER
5920 longbow dr
boulder CO 80301 3299
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key16861058
Report Number1717344-2023-00534
Device Sequence Number12108902
Product Code GEI
UDI-Device Identifier10884521129498
UDI-Public10884521129498
Combination Product (Y/N)N
Initial Reporter CountrySN
PMA/510(K) Number
K110268
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 05/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFORCETRIAD
Device Catalogue NumberFORCETRIAD
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/13/2023
Initial Report FDA Received Date05/03/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE NOTE ON H10
Patient SexUnknown
-
-