SMITH & NEPHEW, INC. 8.0MM FIXED {} ENDCUTTING RMR; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
|
Back to Search Results |
|
| Model Number |
71118224 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/11/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H10: internal complaint reference: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that, during a humerus nail surgery, the procedure began with the reduction of the fracture with the guide and was performed without complications.When determining the diameter of the nail, the surgeon proceeded to pass the distal reamers, starting with the 6 mm, 7mm, and when he passed to the 8.0mm fixed endcutting rmr, when it reached the distal part of the humerus, the reamer split and remained inside the guide.The technique was performed as indicated, the specialist has already worked with the s+n products and knows very well how to manipulate them.The procedure was completed without complications, with a s+n back up device.The incident lasted about 5 minutes and the patient came out in good condition.
|
| |
|
Additional Manufacturer Narrative
|
|
H3, h6: the device was not returned for evaluation.However, the photographs were reviewed, and revealed that the device is fractured.Device batch number was not provided, thus, an evaluation of the manufacturing records could not be performed.A review of complaint history of the previous 12 months revealed similar events for the listed device, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely potential factors that could contribute to the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|
Search Alerts/Recalls
|
|
|