The patient was initially treated for an abdominal aortic aneurysm (aaa) with implant of an afx2 bifurcated stent graft and an afx vela suprarenal proximal extension.Approximately one (1) year post initial procedure, the patient presented at routine follow-up with a type ia endoleak as detected per ct (computed tomography) scan.The physician elected to treat the patient by implanting one (1) endurant (non-endologix) cuff at the proximal portion of the system on (b)(6) 2023, and the endoleak was successfully resolved.The patient was reportedly in stable condition at the conclusion of the re- intervention procedure.
|
|
The reported adverse event/incident was investigated in alignment with endologix operating procedures and work instructions.Where possible, it is endologix practice to make at least three good faith efforts to retrieve a reported adverse event/incident related device as well as medical records and medical imaging.An evaluation of the manufacturing record was completed.A review of the part number/lot number combination needed for device identification shows the device to have been properly manufactured and released in accordance with the device master record.The review confirms there were no manufacturing or processing non-conformities identified that would contribute to the reported adverse event/incident.An evaluation of the device could not be completed.The device was not returned to endologix for evaluation because it remains implanted.A clinical evaluation of the adverse event/incident was completed.An examination of medical records and/or medical imaging received by endologix shows the reported afx2, type ia endoleak and additional endovascular procedure are unconfirmed; it was unclear if the identified endoleak was a type ia or ii endoleak (lumbar) due to the streak artifact from the inferior mesenteric artery and lumbar coiling.This is not consistent with the reported adverse event/incident.Procedure-related harms, device, user, procedure, or anatomy relatedness of this event could not be determined with the medical records available for review.The final patient status was reported as being in stable condition.No additional investigation of this reported adverse event/incident is planned.However, should additional information relevant to the investigation outcome become available, a follow-up report will be submitted.Endologix will continue to monitor this and similar adverse events/incidents.Device iteration is afx with duraply.Corrections: g3 awareness date h6 investigation finding codes; remove 3233 h6 investigation conclusion codes; remove 11.
|