MENTOR TEXAS MENTOR SMOOTH ROUND MODERATE PROFILE; PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE
|
Back to Search Results |
|
| Model Number |
3501610 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Deformity/ Disfigurement (2360)
|
| Date of Event |
01/01/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a 76-year-old female who underwent a breast augmentation primary with a mentor smooth round moderate profile, 125cc saline prosthesis experienced bilateral deflation postoperatively.Patient had an accident with motor vehicle.The issue was diagnosed through physical examination and confirmed with mammogram examination.As a result, patient underwent bilateral removal on (b)(6) 2023.This report relates to the right prosthesis.
|
| |
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: deflation.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The mentor failure analysis lab has received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation completed on may 16 , 2023: the product was returned to mentor for evaluation.Mentor conducted a visual inspection and leak testing of the returned device.Visual analysis of the returned sample revealed that the sal smooth rnd diap 125cc breast implant had a crease/fold extended from the anterior to the posterior view.Leak testing was performed, according to the mentor procedure, and it identified a tear within the crease/fold on the anterior view, measuring approximately 0.2 cm.The evaluation determined that the cause of the deflation could be the trauma experienced.Deflation can occur at any time after implantation, but it is more likely to occur the longer the implant is implanted.Trauma is a known inherent risk of the saline-filled mammary prosthesis.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Capa activity was not required.Manufacturer¿s reference number: (b)(4).
|
| |
|
Search Alerts/Recalls
|
|
|