• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: O&M HALYARD, INC. PFR95 PART FILT RESP; ICP FACIAL PROTECTION PRODUCTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

O&M HALYARD, INC. PFR95 PART FILT RESP; ICP FACIAL PROTECTION PRODUCTS Back to Search Results
Model Number 46727
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 10/16/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Five (5) random samples were inspected.The masks were opened and the elastic head bands were tugged with moderate force.None of the elastic head bands broke or detached from the corner bonds.The elastic remained securely attached inside the corner bonds and the corner bonds remained bonded together.A review of dhr production records was completed.Documented records show all visual and functional inspections were within specifications and there were no nonconformances documented during production of the complaint lot.No root cause was identified.This product incident is documented in the o&m halyard, inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that a o&m halyard, inc.Product is defective or has caused serious injury.
 
Event or Problem Description
The distributor reported that defective headbands tear during use.The following medical device report is being filed retroactive to prior decision that this malfunction would be unlikely to result in a serious injury or illness.A reassessment of this incident has since found that should this malfunction recur in an isolation setting or during direct patient care, the end user or patient may be at an increased risk of contracting a pathogen.For this reason this incident is being reported as a malfunction at this time.Please note that in this incident, the reporter has indicated that no injury or illness resulted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PFR95 PART FILT RESP
Common Device Name
ICP FACIAL PROTECTION PRODUCTS
Manufacturer (Section D)
O&M HALYARD, INC.
1 edison drive
alpharetta GA 30005
Manufacturer (Section G)
LA ADA DE ACUNA
km 4.5 carreterra presa
la amistad
ciudad de acuna, coahuila, mexico 26220
MX   26220
Manufacturer Contact
nichole early
1 edison drive
alpharetta, GA 30005
8287820529
MDR Report Key16865136
Report Number9616096-2020-00013
Device Sequence Number14807745
Product Code MSH
UDI-Device Identifier30680651467271
UDI-Public30680651467271
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K974068
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 05/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number46727
Device Catalogue Number991046727
Device Lot NumberAL0217511
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/16/2020
Initial Report FDA Received Date05/03/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-