• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SURGICAL SPECIALTIES MEXICO S. DE. R.L. DE C.V. STRATAFIX; 2CT-1 #2 PDO 24 X 24

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SURGICAL SPECIALTIES MEXICO S. DE. R.L. DE C.V. STRATAFIX; 2CT-1 #2 PDO 24 X 24 Back to Search Results
Model Number SXPD2B402
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/06/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A review of the device history records for the reported finished good lot and raw material components identified no quality issues during the incoming, manufacturing, in-process, or final inspection processes.No additional complaints have been received for this lot.Sterile samples from the reported lot were not returned for visual review and testing.The actual devices or magnified photos were not returned for review.If samples or photos become available at a later time, they will be reviewed and tested and the results will be included in the file.A follow-up report will be submitted at that time.A potential root cause for a needle detaching when slightly tugged during a procedure could be that the suture was not fully inserted within the end of the needle during the manufacturing process.It¿s also possible the device(s) are being grasped on or near the swaged end, damaging the suture, allowing it to break free from the needle component.However, there are numerous other circumstances that can adversely affect the strength and durability of a suture/needle attachment including but not limited to the misalignment of the suture within the end of the needle, the suture material not placed deep enough within the end of the needle and/or if end is not crimped with enough force to form a secure grip onto the suture material.The ¿precautions¿ section in the ifu for the device states, ¿care should be taken to avoid damage when handling.Avoid crushing or crimping the suture material with surgical instruments such as needle holders and forceps.To avoid damaging needle points and swage areas, grasp the needle in an area one-third (1/3) to one-half (1/2) of the distance from the swaged end to the point¿.Without reviewing the actual reported device(s), receiving magnified photos of the actual devices, or receiving detailed information regarding the method utilized to remove the device from the packaging, preoperative preparation of the device, tools utilized to grasp the devices, details of the procedure performed, or the surgeon¿s technique, a definitive root cause cannot be determined at this time.
 
Event or Problem Description
The health authority reported to our distributor, after the needle passed through the fibroid, it was slightly pulled and there was a problem with the needle pulling free from the suture.The needle remained in the patient's body, and the surgery was immediately suspended.The doctor searched for a long time in the abdominal cavity before finding and removing the needle.Changed to another device to complete the surgery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STRATAFIX
Common Device Name
2CT-1 #2 PDO 24 X 24
Manufacturer (Section D)
SURGICAL SPECIALTIES MEXICO S. DE. R.L. DE C.V.
corredor tijuana rosarito 2000
24702 b, ejido francisco villa
tijuana 22245
MX  22245
Manufacturer (Section G)
SURGICAL SPECIALTIES CORPORATION
247 station drive
suite ne1
westwood MA 02090
Manufacturer Contact
ronald giannangelo
corredor tijuana-rosarito 2000 #24702b
ejido francisco villa
tijuana 22245
MX   22245
MDR Report Key16869187
Report Number3010692967-2023-00018
Device Sequence Number6099982
Product Code NEW
UDI-Device Identifier10705031207882
UDI-Public(01)10705031207882(20)12(17)261130(10)BC28EDN
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K080985
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 05/04/2023
Report Date (Section F) 04/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSXPD2B402
Device Lot NumberBC28EDN
Was the Report Sent to FDA? No
Distributor Facility Aware Date04/06/2023
Device Age30 MO
Event Location Hospital
Date Report to Manufacturer04/27/2023
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/27/2023
Initial Report FDA Received Date05/04/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/28/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age44 YR
Patient SexFemale
-
-