A review of the device history records for the reported finished good lot and raw material components identified no quality issues during the incoming, manufacturing, in-process, or final inspection processes.No additional complaints have been received for this lot.Sterile samples from the reported lot were not returned for visual review and testing.The actual devices or magnified photos were not returned for review.If samples or photos become available at a later time, they will be reviewed and tested and the results will be included in the file.A follow-up report will be submitted at that time.A potential root cause for a needle detaching when slightly tugged during a procedure could be that the suture was not fully inserted within the end of the needle during the manufacturing process.It¿s also possible the device(s) are being grasped on or near the swaged end, damaging the suture, allowing it to break free from the needle component.However, there are numerous other circumstances that can adversely affect the strength and durability of a suture/needle attachment including but not limited to the misalignment of the suture within the end of the needle, the suture material not placed deep enough within the end of the needle and/or if end is not crimped with enough force to form a secure grip onto the suture material.The ¿precautions¿ section in the ifu for the device states, ¿care should be taken to avoid damage when handling.Avoid crushing or crimping the suture material with surgical instruments such as needle holders and forceps.To avoid damaging needle points and swage areas, grasp the needle in an area one-third (1/3) to one-half (1/2) of the distance from the swaged end to the point¿.Without reviewing the actual reported device(s), receiving magnified photos of the actual devices, or receiving detailed information regarding the method utilized to remove the device from the packaging, preoperative preparation of the device, tools utilized to grasp the devices, details of the procedure performed, or the surgeon¿s technique, a definitive root cause cannot be determined at this time.
|
|
The health authority reported to our distributor, after the needle passed through the fibroid, it was slightly pulled and there was a problem with the needle pulling free from the suture.The needle remained in the patient's body, and the surgery was immediately suspended.The doctor searched for a long time in the abdominal cavity before finding and removing the needle.Changed to another device to complete the surgery.
|