• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH SOCKET WRENCH Ø 11.0 MM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH SOCKET WRENCH Ø 11.0 MM Back to Search Results
Model Number 321.15
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Depuy synthese is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthese has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthese or its employees that the report constitutes an admission that the device, depuy synthese, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.D9: complainant part is not expected to be returned for manufacturer review/investigation.E3: reporter is a j&j sales representative.H3, h4, h6: without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that during a revision procedure, a nail was inserted.While inserting screws, the t-handle kept disengaging from the short power shaft, and the inner connecting element broke from malleting.Driver 03.045.003 and retention pin 03.045.004 also stripped/broke in the process of tightening the screw while malleting.After nailing, a va ppfx proximal femur plate was positioned due to the nail.The screw inserted with the cannulated hex driver was stripped due to the torque of the screwdriver.A handle swivel wrench was also used with the articulating tension device; two broke off due to pressure, so a solid one was used instead.There were no patient outcomes or consequences.Concomitant product details: unk - nails (part # unknown, lot # unknown, quantity - unknown), unk - plates: trauma (part # unknown, lot # unknown, quantity - unknown) this report is for one (1) socket wrench ø 11.0 mm this is report 7 of 9 for complaint.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOCKET WRENCH Ø 11.0 MM
Common Device Name
WRENCH
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key16869717
Report Number8030965-2023-05764
Device Sequence Number5460916
Product Code HXC
UDI-Device Identifier10886982190383
UDI-Public(01)10886982190383
Combination Product (Y/N)N
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number321.15
Device Catalogue Number321.15
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/19/2023
Supplement Date Received by Manufacturer05/30/2023
Initial Report FDA Received Date05/04/2023
Supplement Report FDA Received Date06/12/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CANNULATED 2.5MM HEXAGONAL SCREWDRIVER SHAFT.; RETENTION PIN SHORT XL25.; RETENTION PIN W/QUICK-COUPL HEX 12 SHORT.; SCRDRIVER SHORT XL25.; T-HANDLE CANN W/QUICK-COUPL HEX12.; UNK - NAILS.; UNK - PLATES: TRAUMA.; UNK - SCREWS: TRAUMA.
Patient SexUnknown
-
-