| Model Number |
0998-UC-3023-53 |
| Medical Device Problem Code |
Premature Discharge of Battery (1057)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/27/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during preventive maintenance (pm) performed by a getinge service territory manager (stm) cs300 intra-aortic balloon pump (iabp) battery failed run test.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
A supplemental report will be submitted upon completion of our investigation.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected data: d5.
|
| |
|
Additional Manufacturer Narrative
|
|
The getinge field service engineer (fse) that encountered the reported issue during the preventative maintenance (pm) replaced the batteries to fix the issue and performed a full functional and safety checks to meet factory specifications.Unit passed all functional and safety test per factory specifications.The pm was completed and the iabp was then released and cleared for clinical service.
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|