• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMO MANUFACTURING USA, LLC LIQUID OPTICS INTERFACE; OPHTHALMIC FEMTOSECOND LASER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO MANUFACTURING USA, LLC LIQUID OPTICS INTERFACE; OPHTHALMIC FEMTOSECOND LASER Back to Search Results
Model Number 0180-1401
Medical Device Problem Code Decrease in Suction (1146)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/28/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a suction loss occurred during laser fire.No patient injury was reported and the laser procedure was completed successfully.No further detail was provided.
 
Additional Manufacturer Narrative
Section d6a - implant date: not applicable.The interface is not an implantable device.Section d6b - explant date: not applicable.The interface is not an implantable device; therefore, not explanted.Device evaluation: the interface was not returned for evaluation.Therefore, a failure analysis of the complaint device could not be performed.Manufacturing record evaluation: the manufacturing records for the interface were reviewed.The product was manufactured and released according to specification.A search revealed that no additional complaints for this order number have been received.Conclusion: as a result of the investigation, there is no indication of a product quality deficiency.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
Additional information: section h3: device evaluated by manufacturer: no device received in a condition which made analysis impossible.Rga 204048097 was received at jjsv irvine on 05/11/2023.The contents of the package were received loose within a bag.Individual liquid optics interface (loi) components could not be associated with their corresponding complaint events or lot numbers.Product testing cannot be performed.Please note that the information reported in sections d2 (common device name), h.6 (health effect -clinical code and medical device problem code) and section g.1 (manufacturer contact e-mail) is information that remains unchanged from the initial emdr report.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIQUID OPTICS INTERFACE
Common Device Name
OPHTHALMIC FEMTOSECOND LASER
Manufacturer (Section D)
AMO MANUFACTURING USA, LLC
510 cottonwood drive
milpitas CA 95035
Manufacturer Contact
somyata nagpal
1700 e st andrew place
santa ana, CA 92705
7142478552
MDR Report Key16871646
Report Number3012236936-2023-01111
Device Sequence Number5478231
Product Code OOE
UDI-Device Identifier05050474609761
UDI-Public(01)05050474609761(17)230623(10)19975711
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K170322
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 06/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/23/2023
Device Model Number0180-1401
Device Catalogue Number0180-1401
Device Lot Number19975711
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/10/2023
Supplement Date Received by Manufacturer06/06/2023
Initial Report FDA Received Date05/04/2023
Supplement Report FDA Received Date06/09/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/24/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CATALYS S/N: (B)(6).
Patient SexUnknown
-
-