• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS Back to Search Results
Model Number VTICM5 13.2
Medical Device Problem Codes Device Dislodged or Dislocated (2923); Patient Device Interaction Problem (4001)
Health Effect - Clinical Codes Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/12/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Claim# (b)(4).
 
Event or Problem Description
The reporter indicated that the surgeon implanted an implanted collamer lens into the patient's eye on (b)(6) 2023.It was indicated that the lens was displaced through the pupil.Lens remains implanted.Attempts to obtain additional information have not been successful.If additional information is received a supplemental medwatch report will be submitted.
 
Additional Manufacturer Narrative
Additional information: h6: health effect- clinical code: 4581- "iris atrophy." b5: the reporter indicated that the surgeon implanted a 13.2mm vticm5_13.2 implantable collamer lens of diopter -9.50/3.0/141 (sphere/cylinder/axis) into the patient's right eye (od) on (b)(6) 2023.The patient experienced lens dislocation and iris atrophy.Reportedly "half of the lens displaced through the pupil" and "planning to reposition".The lens was repositioned and this resolved the problem.Status of the eye: "back to normal".The cause of the event: patient related factor - "patient inadvertently rubbed eye".Claim#: (b)(4).
 
Additional Manufacturer Narrative
Corrected data: h6-health effect- impact code: "4628" should be added.H6- type of investigation code: 4110- lens work order search-no similar complaint event(s) within associated lots were found.Claim# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Common Device Name
PHAKIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SUGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
joselene enriquez
1911 walker avenue
monrovia, CA 91016
MDR Report Key16873135
Report Number2023826-2023-01664
Device Sequence Number20188610
Product Code MTA
UDI-Device Identifier00841542116541
UDI-Public00841542116541
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/12/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2024
Device Model NumberVTICM5 13.2
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/12/2023
Supplement Date Received by Manufacturer05/23/2023
06/14/2023
Initial Report FDA Received Date05/04/2023
Supplement Report FDA Received Date05/23/2023
06/14/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age26 YR
Patient SexMale
-
-