• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MMJ SA DE CV (USD) SHILEY; TUBE TRACHEOSTOMY AND TUBE CUFF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MMJ SA DE CV (USD) SHILEY; TUBE TRACHEOSTOMY AND TUBE CUFF Back to Search Results
Model Number 4CFN
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Codes Pain (1994); Discomfort (2330); Unintended Extubation (4564)
Date of Event 04/19/2023
Type of Reportable Event Serious Injury
Event or Problem Description
According to the reporter, the length of the trach has been inconsistent with some were very short and long.The inner cannulas were often jagged at the edge and appeared not to be properly cut.Often the inner cannulas would stick out past the outer trach shaft.Some trachs had tiny threads or shavings of plastic sticking out from the outer edge of the outer trach shaft.The trachs did not seem to fit right and was causing a great deal of pain and discomfort to patient's airway and around the stoma.The patient had difficulty with breathing off and on when the pain got worse.The trach was changed frequently for relief from great irritation on the patient's chest.
 
Additional Manufacturer Narrative
D10 concomitant product/s: 4cfn, 4cfn, 4 fen, trach cann w reus ic x1 lot #:22c0832jzx.4cfn, 4cfn, 4 fen, trach cann w reus ic x1 lot #:22c0832jzx.4cfn, 4cfn, 4 fen, trach cann w reus ic x1 lot #:22c0832jzx.4cfn, 4cfn, 4 fen, trach cann w reus ic x1 lot #:22c0832jzx.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHILEY
Common Device Name
TUBE TRACHEOSTOMY AND TUBE CUFF
Manufacturer (Section D)
MMJ SA DE CV (USD)
ave henequen no 1181 desarroll
ciudad juarez 32590
MX  32590
Manufacturer (Section G)
MMJ SA DE CV (USD)
ave henequen no 1181 desarroll
ciudad juarez 32590
MX   32590
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key16876377
Report Number2936999-2023-00384
Device Sequence Number12191974
Product Code JOH
UDI-Device Identifier50884522006918
UDI-Public50884522006918
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K962173
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer,Health Professional
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 05/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number4CFN
Device Catalogue Number4CFN
Device Lot Number22C0832JZX
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/19/2023
Initial Report FDA Received Date05/05/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age53 YR
Patient SexFemale
-
-