• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL INQUIRY¿ STEERABLE DIAGNOSTIC CATHETER; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL INQUIRY¿ STEERABLE DIAGNOSTIC CATHETER; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING Back to Search Results
Model Number IBI-81104
Medical Device Problem Codes Difficult to Remove (1528); Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/02/2023
Type of Reportable Event Serious Injury
Event or Problem Description
During a ventricular tachycardia (vt) procedure, the catheter could not be removed and an additional puncture site was needed.Upon insertion into the right atrium, the catheter kinked while maneuvering and could not be straightened.An attempt to extract the catheter was not possible, as it did not go through the sheath.The decision was made to leave the catheter in the vein and to do an additional puncture to place another inquiry catheter through a new sheath.The first catheter was left in the veins throughout the procedure.A second catheter was placed via an additional puncture without problems.Following the procedure, all catheters were removed with no consequences to the patient.
 
Additional Manufacturer Narrative
One decapolar, inquiry steerable diagnostic catheter was received for evaluation.A bend in the catheter shaft was noted 1.3¿ and the distal shaft deflected in an 's' shape when actuating the steering mechanism.The catheter was unable to return to a neutral position when straightened due to the bend in the catheter shaft.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause for the additional puncture site was due to removal difficulties related to a bent shaft and s curve.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INQUIRY¿ STEERABLE DIAGNOSTIC CATHETER
Common Device Name
CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
Manufacturer (Section D)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
Manufacturer (Section G)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS   1897-4050
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key16876850
Report Number3008452825-2023-00191
Device Sequence Number17964716
Product Code DRF
UDI-Device Identifier05414734302975
UDI-Public05414734302975
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K961924
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/28/2026
Device Model NumberIBI-81104
Device Catalogue NumberIBI-81104
Device Lot Number8921032
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/02/2023
Supplement Date Received by Manufacturer06/23/2023
Initial Report FDA Received Date05/05/2023
Supplement Report FDA Received Date06/27/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/02/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-