Model Number PLC271200
Medical Device Problem Code Patient Device Interaction Problem (4001)
Health Effect - Clinical Code Clinical Endoleak (4843)
Date of Event 04/10/2023
Type of Reportable Event Serious Injury
Event or Problem Description
On (b)(6) 2018, the patient was treated for an unknown issue.The physician implanted a gore® excluder® aaa endoprosthesis featuring c3® delivery system and gore® excluder® aaa endoprosthesis.The patient tolerated the procedure.On (b)(6) 2023, a type ib endoleak was detected.The physician plans to do a reintervention.Revision was done on move medical case (b)(4).The endoleak was ballooned, and patient tolerated the procedure.
 
Additional Manufacturer Narrative
H6: code c19 - the review of the manufacturing paperwork verified that this lot met all pre-release specifications.H6: code d12 ¿ according to the gore® excluder® aaa endoprosthesis instructions for use, adverse events that may occur and/or require intervention include, but are not limited to, endoleaks.An evaluation of returned images was performed: imaging evaluation: summary: the imaging evaluation performed by a clinical imaging specialist showed the following: one time point is available for evaluation: post-implantation cta dated (b)(6) 2023.Maximum diameter in the left common iliac artery is 45mm.There is lack of distal, contralateral leg (plc271200), device apposition.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® EXCLUDER® AAA ENDOPROSTHESIS
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL PHOENIX 1 B/P
32360 n. north valley parkway
phoenix AZ 85085
Manufacturer Contact
greg rawlings
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key16877462
Report Number3007284313-2023-02481
Device Sequence Number20291053
Product Code MIH
UDI-Device Identifier00733132618545
UDI-Public00733132618545
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
P020004
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/01/2021
Device Model NumberPLC271200
Device Catalogue NumberPLC271200
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/10/2023
Initial Report FDA Received Date05/05/2023
Date Device Manufactured04/02/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age65 YR
Patient SexMale
-
-