• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH MASIMO SEDLINE MODULE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH MASIMO SEDLINE MODULE Back to Search Results
Model Number 867186
Medical Device Problem Codes Increase in Pressure (1491); High Readings (2459)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/09/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A follow-up report will be submitted upon completion of the investigation.Reporting institution name: (b)(6) hospital.Reporting address state: (b)(6).
 
Event or Problem Description
The sedline device is a depth of anaesthesia monitor.It analyses the eeg of the anaesthetised patient.This analysis is displayed as a psi number.The range 25-50 is suggested to show adequate surgical anaesthesia.The raw eeg is also displayed, as is the sef - spectral edge frequency - another indicator of adequate anaesthesia.The use of depth of anaesthesia devices is recommended with total intravenous anaesthesia to prevent awareness during surgery.They are also considered useful to prevent overdosage of anaesthetic agents, and thus minimise the chance of post operative cognitive dysfunction.The psi displayed by sedline is proving unreliable.The number spontaneously climbs, suggesting inadequate anaesthesia, whilst all other clinical signs remain normal.
 
Additional Manufacturer Narrative
H6 device problem code was corrected.
 
Additional Manufacturer Narrative
A philips remote service engineer (rse) attempted to reach out to the customer and received no response.It is unknown if the device was evaluated or what caused the issue.Based on the information available and the testing conducted, the cause of the reported problem is unknown.The reported problem was not confirmed.It is unknown if the device is back in service due to insufficient information.The customer did not call back or email back for further support.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MASIMO SEDLINE MODULE
Common Device Name
MASIMO SEDLINE MODULE
Manufacturer (Section D)
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM  71034
Manufacturer (Section G)
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM   71034
Manufacturer Contact
hauke schik
3000 minuteman rd
andover, MA 01810
6172455900
MDR Report Key16880005
Report Number9610816-2023-00208
Device Sequence Number20291152
Product Code OLW
UDI-Device Identifier00884838051515
UDI-Public00884838051515
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K003038
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number867186
Device Catalogue Number867186
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/21/2023
Supplement Date Received by Manufacturer04/21/2023
07/27/2023
Initial Report FDA Received Date05/05/2023
Supplement Report FDA Received Date06/14/2023
08/02/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/02/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-