| Model Number |
WA62820 |
| Medical Device Problem Code |
Crack (1135)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/12/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported to olympus, that during manual cleaning of a6282 used in surgical laparoscopic surgery.A plastic part in front of the electrode chipped.There was no patient injury or adverse event reported to olympus.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for evaluation and repair.The investigation is still in progress.Therefore, the root cause of the reported defect cannot be determined at this time.However, if additional information becomes available, this report will be supplemented accordingly.In general, the customer is required to inspect the device for defects.Check the function of all devices.And have alternate equipment prior to use.Olympus will continue to monitor the field performance of this device.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and correction to g2 (also selected ¿other¿ which was inadvertently left off the initial report).The customer's complaint was confirmed.There are scratches on the tip of the electrode, chips and cracks in the tip.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely that the event occurred due to improper handling/impaction to the tip.Olympus will continue to monitor field performance for this device.
|
| |
|