• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH MASIMO SEDLINE MODULE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH MASIMO SEDLINE MODULE Back to Search Results
Model Number 867186
Medical Device Problem Codes Increase in Pressure (1491); High Readings (2459)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/09/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A follow-up report will be submitted upon completion of the investigation.Reporting institution name: (b)(6) hospital.Reporting address state: (b)(6).
 
Event or Problem Description
The sedline device is a depth of anesthesia monitor.It analyses the eeg of the anaesthetized patient.This analysis is displayed as a psi number.The range 25-50 is suggested to show adequate surgical anesthesia.The raw eeg is also displayed, as is the sef - spectral edge frequency - another indicator of adequate anesthesia.The use of depth of anesthesia devices is recommended with total intravenous anesthesia to prevent awareness during surgery.They are also considered useful to prevent overdosage of anesthetic agents, and thus minimize the chance of post operative cognitive dysfunction.The psi displayed by sedline is proving unreliable.The number spontaneously climbs, suggesting inadequate anesthesia, whilst all other clinical signs remain normal.The device was in clinical use at the time the issue was discovered.There was no adverse event or patient harm reported.
 
Additional Manufacturer Narrative
A philips remote service engineer (rse) attempted to reach out to the customer and received no response.It is unknown if the device was evaluated or what caused the issue.Based on the information available and the testing conducted, the cause of the reported problem is unknown.The reported problem was not confirmed.It is unknown if the device is back in service due to insufficient information.The customer did not call back or email back for further support.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.H3 other text : no response from customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MASIMO SEDLINE MODULE
Common Device Name
MASIMO SEDLINE MODULE
Manufacturer (Section D)
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM  71034
Manufacturer (Section G)
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM   71034
Manufacturer Contact
hauke schik
3000 minuteman rd
andover, MA 01810
6172455900
MDR Report Key16885349
Report Number9610816-2023-00213
Device Sequence Number14963315
Product Code OLW
UDI-Device Identifier00884838051515
UDI-Public00884838051515
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K003038
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number867186
Device Catalogue Number867186
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/21/2023
Supplement Date Received by Manufacturer04/21/2023
07/27/2023
Initial Report FDA Received Date05/08/2023
Supplement Report FDA Received Date06/14/2023
08/02/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/02/2022
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-