| Model Number |
10973HD |
| Medical Device Problem Codes |
Device Handling Problem (3265); Sharp Edges (4013)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/10/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation on-going.Product currently will stay at the facility.Should the product be sent in later with additional information/investigation results a supplemental medwatch will be submitted unsolicited.The event is filed under internal karl storz complaint id (b)(4).As of today, one other similar case has been found.The manufacture continues to review and investigate.
|
| |
|
Event or Problem Description
|
|
It was reported that there was event with a 10973hd video mediastinoscope.The surgeon complained that this hd version felt to sharp at the tip and complained about the angle at the distal portion of the scope.Surgeon felt there was a real risk of perforating the trachea.No patient injury, patient under anesthesia, but no incision made.
|
| |
|
Additional Manufacturer Narrative
|
|
"health effect- impact code: appropriate term/code not available-surgery aborted".
|
| |
|
Additional Manufacturer Narrative
|
|
Per manufacture findings: after a thorough examination of the mediastinoscope in our possession, the customer's information regarding the sharp tip / sharp working end cannot be confirmed.The working end is clearly blunt and evenly rounded.The imaging is clear and distinct with no discernible image impairments.The connector plug as well as the connecting cable are without defects.The customer's complaint cannot be substantiated.No production defect was found in the product.The mediastinoscope complies with the currently valid specifications the internal karlstorz complaint number is (b)(4).
|
| |
|