• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARD® INLAY OPTIMA® URETERAL STENT WITH HYDROGLIDE¿ GUIDEWIRE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARD® INLAY OPTIMA® URETERAL STENT WITH HYDROGLIDE¿ GUIDEWIRE Back to Search Results
Model Number 787830
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/17/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a physician stated in an online survey that the patient experienced stone formation, extravasation, fistula formation, pain or discomfort, hydronephrosis prior to the device placement and stone formation, ureteral reflux, stent dislodgement complications were directly attributable to the device and ureteral stricture was needed for stone formation in relation to inlay optima ureteral stent with hydroglide guidewire, 8.0 fr.X 30cm.
 
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The device was not returned.
 
Event or Problem Description
It was reported that a physician stated in an online survey that the patient experienced stone formation, extravasation, fistula formation, pain or discomfort, hydronephrosis prior to the device placement and stone formation, ureteral reflux, stent dislodgement complications were directly attributable to the device and ureteral stricture was needed for stone formation in relation to inlay optima ureteral stent with hydroglide guidewire, 8.0 fr.X 30cm.
 
Additional Manufacturer Narrative
The reported event is inconclusive because no sample was returned for evaluation and further investigation was not conclusive.Though a specific cause cannot be determined, a potential root cause for this event could be material selection or part geometry.The device was used for treatment purposes.It was unknown if the device had met all relevant specifications or resulted in the reported failure.A dhr review cannot be completed as the lot number is unknown.The instructions for use were found adequate and state the following: "with any ureteral stent, migration is a possible complication, which could require medical intervention for removal.Selection of too short a stent may result in migration." "the insertion of a ureteral stent should only be done by those individuals who have comprehensive training in the techniques and risks of the procedure." h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® INLAY OPTIMA® URETERAL STENT WITH HYDROGLIDE¿ GUIDEWIRE
Common Device Name
URETERAL STENT
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key16886038
Report Number1018233-2023-03235
Device Sequence Number6326722
Product Code FAD
UDI-Device Identifier10801741015622
UDI-Public(01)10801741015622
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K043193
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Study
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number787830
Device Catalogue Number787830
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/17/2023
Supplement Date Received by Manufacturer07/07/2023
Initial Report FDA Received Date05/08/2023
Supplement Report FDA Received Date07/18/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;