| Model Number |
IPN041720 |
| Medical Device Problem Code |
Gas/Air Leak (2946)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/10/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Qn# (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that "the doctor found cuff leakage when using on patient.Then changed new one, no impact on patient." the patient status is reported as "fine".
|
| |
|
Event or Problem Description
|
|
It was reported that "the doctor found cuff leakage when using on patient.Then changed new one, no impact on patient." the patient status is reported as "fine".
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).The actual sample involved in the complaint was not returned for evaluation; therefore, a representative sample was taken from current production at the manufacturing facility.A visual exam was performed on the representative sample and no defects were observed.The blue cuff and the clear cuff were then inflated to 25mbar using a calibrated endotest.The cuffs were inflated and deflated with no issues.The sample was then immersed in water to check for leaks.No bubbles were observed coming from the cuffs indicating there were no leaks.A device history record (dhr) review was performed and there were no issues found that could relate to the reported complaint.Without the device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
|
| |
|