H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one hickman s/l catheter was returned for evaluation.Functional, tactile, gross visual and microscopic visual evaluations were performed.The investigation is confirmed for the reported cuff bonding issue as the tissue cuff was intact and had residue and it was noted to move freely on the catheter.Furthermore, lack of adhesive was noted on the tissue cuff area in the catheter.The definitive root cause could not be determined based upon available information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: b5, d4 (expiration date: 01/2026), h11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|