| Catalog Number |
05.001.204 |
| Medical Device Problem Codes |
Fire (1245); Complete Loss of Power (4015)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported from switzerland that during service and evaluation, it was determined that the battery charger device would not run ¿ power supply was damaged, had a blown fused, spark/fire, and did not function.It was further determined that the device failed pretest for check firmware revision and power-on test.It was noted in the service order that the device had no power.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention, or prolonged hospitalization.The exact date of the event was not reported.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.The actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the battery charger device had spark/fire and would not run.Therefore, the reported condition was confirmed.The assignable root cause of this condition was determined to be traced to component failure due to wear.Udi: (b)(4).
|
| |
|
Search Alerts/Recalls
|
|
|