| Model Number |
21-4590-24 |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Radiation Leak (1357); Infusion or Flow Problem (2964)
|
| Health Effect - Clinical Code |
Swelling/ Edema (4577)
|
| Date of Event |
04/23/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient was admitted to the hospital due to fourth-degree myelosuppression.During the infusion of nutrient solution, the infusion "suddenly stopped", and the patient's neck was found to be swollen during the process.The catheter was confirmed to be broken and cracked via digital subtraction angiography (dsa) at the radiology department.Leakage of contrast medium from the catheter.No adverse effects reported.
|
| |
|
Additional Manufacturer Narrative
|
|
One device was returned for investigation.Visual inspection occurred and a section of the catheter was found to be attached to the port with no scuffs, pinch marks, cracks, crazing cuts or other damages that could contribute to the failure mode.A leak test was performed and the device showed no leaks were present.The reported complaint is not confirmed.Root cause cannot be associated with manufacturing process since the failure could not be reproduced.No corrective actions are required.A review of the device history records shows there were no observations recorded during manufacture to suggest an issue of this nature would occur with this lot of products.
|
| |
|