MAUDE Adverse Event Report: ABBOTT MEDICAL REGENT HEART VALVE WITH FLEX CUFF; HEART-VALVE, MECHANICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL REGENT HEART VALVE WITH FLEX CUFF; HEART-VALVE, MECHANICAL Back to Search Results
Model Number 23AGFN-756
Medical Device Problem Code Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/12/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported that on (b)(6) 2023, a 23mm sjm regent heart valve w/ flex cuff was selected for an implant.The leaflets were moving normally when tested at the time of device preparation.Plastic surgical instruments and physiological saline were used to test leaflet function.The device was successfully implanted in a patient.Post procedure, it was found that the valve leaflet did not open and close properly and was explanted.The leaflet was tested after explant but did not moved normally.On the same day, a new 23mm sjm regent heart valve w/ flex cuff was implanted as a replacement.The patient took warfarin after procedure and latest inr is 2.5.The patient is recovering.
 
Additional Manufacturer Narrative
The reported event of valve leaflets not opening and closing normally after implant could not be confirmed.Investigation found no anomalies on the valve leaflets, and functional testing at the time of manufacturing and upon return to abbott indicated the valve functioned normally.This test ensures proper leaflet coaptation and hemodynamic performance.Information from the field indicated that there was resistance or sticking while opening/closing the leaflets after the valve was sutured during the procedure.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.The cause of the reported incident could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REGENT HEART VALVE WITH FLEX CUFF
Common Device Name
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL PUERTO RICO, INC. REG#2648612
20 b st caguas west park
caguas 00725
*   00725
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key16890484
Report Number2135147-2023-01991
Device Sequence Number6173542
Product Code LWQ
UDI-Device Identifier05414734006095
UDI-Public05414734006095
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P810002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/10/2025
Device Model Number23AGFN-756
Device Catalogue Number23AGFN-756
Device Lot Number7736555
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/04/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/13/2023
Supplement Date Received by Manufacturer06/21/2023
Initial Report FDA Received Date05/08/2023
Supplement Report FDA Received Date07/14/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/11/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age61 YR
Patient SexMale
Patient Weight65 KG
-
-