MEDACTA INTERNATIONAL SA MPACT ACETABULAR SHELL Ø52 MULTI-HOLE; HIP ACETABULAR SHELL Back to Search Results
Model Number 01.32.152MH
Medical Device Problem Code Failure to Osseointegrate (1863)
Health Effect - Clinical Code Inadequate Osseointegration (2646)
Date of Event 04/14/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Batch review performed on (b)(6)2023: lot 2106973: (b)(4) items manufactured and released on 22-nov-2021.Expiration date: 2026-11-04.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold with no similar reported case during the period of review.Clinical evaluation performed by meadacta medical affairs manager: revision 1 year after the primary tha surgery due to a loose cup.The patient came in reporting pain.The cause of the loosening is unknown.The radiographic images lead us to suspect that the medial acetabular border could have been violated during cup insertion, therefore limiting the possibility of the cup to achieve stable steointegration.The surgeon revised successfully the cup, head and liner.
 
Event or Problem Description
At about 1 year after the primary surgery, the patient came in reporting pain due to a loose cup and the cause is unknown.The surgeon revised successfully the cup, head and liner.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MPACT ACETABULAR SHELL Ø52 MULTI-HOLE
Common Device Name
HIP ACETABULAR SHELL
Manufacturer (Section D)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, switzerland 6874
SZ  6874
Manufacturer (Section G)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, switzerland 6874
SZ   6874
Manufacturer Contact
marco giannessi
strada regina
castel san pietro, switzerland 6874
SZ   6874
MDR Report Key16893410
Report Number3005180920-2023-00344
Device Sequence Number6162762
Product Code LPH
UDI-Device Identifier07630030810800
UDI-Public07630030810800
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K132879
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 05/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number01.32.152MH
Device Catalogue Number01.32.152MH
Device Lot Number2106973
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/14/2023
Initial Report FDA Received Date05/09/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/22/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-