| Model Number |
01.32.152MH |
| Medical Device Problem Code |
Failure to Osseointegrate (1863)
|
| Health Effect - Clinical Code |
Inadequate Osseointegration (2646)
|
| Date of Event |
04/14/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Batch review performed on (b)(6)2023: lot 2106973: (b)(4) items manufactured and released on 22-nov-2021.Expiration date: 2026-11-04.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold with no similar reported case during the period of review.Clinical evaluation performed by meadacta medical affairs manager: revision 1 year after the primary tha surgery due to a loose cup.The patient came in reporting pain.The cause of the loosening is unknown.The radiographic images lead us to suspect that the medial acetabular border could have been violated during cup insertion, therefore limiting the possibility of the cup to achieve stable steointegration.The surgeon revised successfully the cup, head and liner.
|
| |
|
Event or Problem Description
|
|
At about 1 year after the primary surgery, the patient came in reporting pain due to a loose cup and the cause is unknown.The surgeon revised successfully the cup, head and liner.
|
| |
|