| Model Number |
VERSION 1.0 |
| Medical Device Problem Codes |
Detachment of Device or Device Component (2907); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/13/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during a recanalization procedure for thrombosed stent in common femoral via contralateral groin access, the tip of the catheter allegedly detached from the catheter once the device was started running in a non-calcified segment for approximately 3 seconds.It was further reported that catheter still maintained barely attached.The procedure was completed by using another device.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: the medical device manufacturer (d3) and manufacturing location (g1) for the straub product was selected as unknown due to system limitations.The correct medical device manufacturer and manufacturing location are straub medical us.H10: as the lot number for the device was provided, a review of the device history records will be performed.The return of the sample is pending.The investigation of the reported event is currently underway.H10: d4 (expiry date: 07/2024).H11: section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device pending return.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: the medical device manufacturer (d3) and manufacturing location (g1) for the straub product was selected as unknown due to system limitations.The correct medical device manufacturer and manufacturing location are straub medical us.H10: manufacturing review: a manufacturing review was conducted and there was nothing found to indicate there was a manufacturing-related cause for this event.Investigation summary: the sample was returned for evaluation.A physical investigation was performed for the catheter.During physical investigation a tube damage was noted at 20 cm from the tip of the catheter.It was also observed helix had a slight kink inside of the handle window.Due to the high amount of coagulated material inside of the catheter, it was not able to aspirate.Therefore, the investigation is confirmed for the reported partial detachment of the rotarex catheter.A definitive root cause could not be determined based upon the available information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 07/2024).H11: section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that during a recanalization procedure for thrombosed stent in common femoral via contralateral groin access, the tip of the catheter allegedly detached from the catheter once the device was started running in a non-calcified segment for approximately 3 seconds.It was further reported that catheter still maintained barely attached.The procedure was completed by using another device.There was no reported patient injury.
|
| |
|