| Catalog Number |
03003523001 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/22/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The field service engineer found the chloride electrode needed replacement.He checked the ise system and observed no issues, performed an ise performance check and all passed, and replaced the mixing tower.The customer changed the electrode and ran qc with results within the specified ranges.The investigation determined the service actions resolved the issue.H3 other text : na.
|
| |
|
Event or Problem Description
|
|
There was an allegation of questionable ise indirect gen chloride results from cobas integra 400 plus analyzer serial number (b)(6).Sample 1 initial result was 141 mmol/l and was reported outside of the laboratory.The doctor called questioning the result.The repeat result was 109 mmol/l.On (b)(6) 2023, sample 2 initial result was 136 mmol/l and the repeat result was 108.The questionable result for this sample was not reported outside of the laboratory.The repeat results were believed correct.
|
| |
|